loading img

No.8152-50

- Discontinued - Digital Panel Meter

SK-EM-01

JAN code : 4974425815258

Calibration : May be available depends on probes

JCSS Accredited Calibration : May be available depends on probes

This product is discontinued.

Digital Panel Meter SK-EM-20 is the inherited model.

    Feature

    Bright, highly visible LCD with a digit as tall as 20 mm

    Thermocouples, thermistor, and analog voltage input are available

    Compact size of 96 x 48 mm applies to DIN standards

    Ideal for Temperature/Humidity Transmitter SK-RHC series and Infrared Thermometer SK-8200 series

    Conforms to RoHS

    Description

    Measuring ranges and codes

    Input method Code Range (°C) Range (°F) Remarks
    Thermo-
    couple
    B 01 0 to 1800 0 to 3300 *2
    R 02 0 to 1700 0 to 3100
    S 03 0 to 1700 0 to 3100
    K 04 −199.9 to 800.0 −300 to 1500
    05 0 to 1200 0 to 2200
    E 06 0 to 700 0 to 1300
    J 07 0 to 600 0 to 1100
    T 08 −199.9 to 300.0 −300 to 600 *3
    N 09 0 to 1300 0 to 2300
    U 10 −199.9 to 300.0 −300 to 600 *3
    L 11 0 to 600 0 to 1100
    WRe5-26 12 0 to 2300 0 to 4200
    Therm-
    istor
    Pt 31 −199.9 to 600.0 −300 to 1100 *4
    32 −100.0 to 100.0 −150.0 to 200.0
    JPt 33 −199.9 to 500.0 −300 to 1000 *4
    34 −100.0 to 100.0 −150.0 to 200.0
    Analog
    voltage
    0–10 mV 71 Scaling to 0.0 to 100.0 available
    Scaling supported range: −1999 to 9999 Unit
    Span: 10 to 10000 Unit
    0–5 V 81
    1–5 V 82
    0–10 V 83

    *1 Factory setting: type-K thermocouple (code 05, 0 to 1200°C)

    *2 Accuracy of type-B thermocouple below 400°C: ±5% F.S.

    *3 Accuracy of type-T and U thermocouple: ±0.5% F.S. (−100 to 0°C), ±1% F.S. (< −100°C)

    *4 Pt (code 31) and JPt (code 33) scale out at temperature below −240.0°C (−400°F)

    Specifications

    Cat. No. 8152-50
    Product name Digital Panel Meter
    Model name SK-EM-01
    Measuring range See "Measuring ranges and codes"
    Display accuracy ±5% F.S. (type-B thermocouple below 400°C)
    ±0.5% F.S. (type-T and U thermocouples −100 to 0°C)
    ±1% F.S. (T and U below −100°C)
    ± (0.3% F.S. + 1 digit) (other)
    Resolution Depends on probes
    Waterproofness IP66 dustproof/waterproof
    Operating ambient −10 to 50°C, less than 90%rh without condensing, altitude below 2000 m
    Complied regulations IEC61010-1, EN61010-1, EMC: EN61326-1
    Dimensions 96 (W) x 48 (H) x 111 (D) mm with depth behind the panel 100 mm
    Weight 250 g
    Power 100&ndash240 VAC