    Feature

    Bright, highly visible LCD with a digit as tall as 20 mm

    Thermocouples, thermistor, and analog voltage input are available

    Compact size of 96 x 48 mm applies to DIN standards

    Ideal for Temperature/Humidity Transmitter SK-RHC series and Infrared Thermometer SK-8200 series

    Conforms to RoHS

    Description

    Measuring ranges and codes

    Input methodCodeRange (°C)Range (°F)Remarks
    Thermo-
    couple
    B010 to 18000 to 3300*2
    R020 to 17000 to 3100
    S030 to 17000 to 3100
    K04−199.9 to 800.0−300 to 1500
    050 to 12000 to 2200
    E060 to 7000 to 1300
    J070 to 6000 to 1100
    T08−199.9 to 300.0−300 to 600*3
    N090 to 13000 to 2300
    U10−199.9 to 300.0−300 to 600*3
    L110 to 6000 to 1100
    WRe5-26120 to 23000 to 4200
    Therm-
    istor
    Pt31−199.9 to 600.0−300 to 1100*4
    32−100.0 to 100.0−150.0 to 200.0
    JPt33−199.9 to 500.0−300 to 1000*4
    34−100.0 to 100.0−150.0 to 200.0
    Analog
    voltage
    0–10 mV71Scaling to 0.0 to 100.0 available
    Scaling supported range: −1999 to 9999 Unit
    Span: 10 to 10000 Unit
    0–5 V81
    1–5 V82
    0–10 V83

    *1 Factory setting: type-K thermocouple (code 05, 0 to 1200°C)

    *2 Accuracy of type-B thermocouple below 400°C: ±5% F.S.

    *3 Accuracy of type-T and U thermocouple: ±0.5% F.S. (−100 to 0°C), ±1% F.S. (< −100°C)

    *4 Pt (code 31) and JPt (code 33) scale out at temperature below −240.0°C (−400°F)

    Specifications

    Cat. No.8152-50
    Product nameDigital Panel Meter
    Model nameSK-EM-01
    Measuring rangeSee "Measuring ranges and codes"
    Display accuracy±5% F.S. (type-B thermocouple below 400°C)
    ±0.5% F.S. (type-T and U thermocouples −100 to 0°C)
    ±1% F.S. (T and U below −100°C)
    ± (0.3% F.S. + 1 digit) (other)
    ResolutionDepends on probes
    WaterproofnessIP66 dustproof/waterproof
    Operating ambient−10 to 50°C, less than 90%rh without condensing, altitude below 2000 m
    Complied regulationsIEC61010-1, EN61010-1, EMC: EN61326-1
    Dimensions96 (W) x 48 (H) x 111 (D) mm with depth behind the panel 100 mm
    Weight250 g
    Power100&ndash240 VAC