SK-8940 is an infrared thermometer with a wide measuring range from −40°C to 500°C. Distance-to-spot ratio is 10:1 thus it enables measurement of a narrower area. Emissivity is adjustable between 0.30 and 1.00 in steps of 0.01. Calibration traceability certificate service is available.

    Feature

    ■ One of the best advantages of a contactless infrared thermometer is its quick response. With SK-8940 you can easily and instantly measure surface temperature in one hand.

    ■ Applicable for a wide range of temperature from −40 to 500°C. Using a single-point laser marker helps you identify the target at a glance.

    ■ D:S = 10:1 (D is distance to the object, S is a diameter of a spot circle)

    Emissivity is adjustable in steps of 0.01.

    ■ Certified for PSC, a standard based on Japan's Consumer Product Safety Act, as a class II portable laser device.

    ■ LCD backlight improves readability in a dark place.

    ■ Turns off the power automatically if the keys are not pressed for 10 seconds, to prevent from running out the battery even when you forget to turn it off.

    ■ Automatic hold function keeps the measured value on the LCD for 10 seconds.

    ■ Traceable calibration available

    Description

    ■ Caution! This device is for industrial use, not for clinical use.

    ■ What is the distance-to-spot ratio?

    A distance-to-spot ratio, which is the ratio of the Distance between the object and the thermometer to the size of the Spot, is preset at D:S = 10:1 for SK-8940. For example, if the distance is 500 mm, a spot circle of diameter 50 mm can be measured. Make sure that the object is larger than the circle.

    The spot circle is defined as an area to receive 90% of the emitted energy from the object.

    ■ Warning: Do not look into the laser marker source or point it on eyes as it may damage eyes.

    SK-8940 conforms to the Class II in the Standard for Safe Use of Laser Products (JIS C6802) which is defined as follows: A visible light (wavelength between 400 and 700 nm) with an output level less than 1 mW that a human's protective response can resist.

    PSC mark is labeled on the device to prove its quality based on Japan's Consumer Product Safety Act.

    Specifications

    Cat. No. 8266-00
    Product name Infrared Thermometer
    Model name SK-8940
    Measuring range −40 to 500°C
    Accuracy ±2%rdg or ±2°C, whichever is larger (−19.9 to 450°C)
    ±4%rdg or ±4°C, whichever is larger (other)
    under blackbody calibration with ambient temperature 23±5°C, emissivity 0.95
    Repeatability ±0.1°C
    Resolution 0.1°C (below 100°C)
    1°C (100°C and up)
    Emissivity Adjustable between 0.30 and 1.00 in steps of 0.01
    Distance coefficient D:S = 10:1 (where D is distance, S is diameter of spot circle)y
    Detector Thermopile
    Measured wavelength 8 to 14 μm
    Laser marker Source wavelength: 650 nm
    Output: Less than 1 mW
    Certified for PSC, a standard based on Japan's Consumer Product Safety Act, as a class II portable laser device.
    Operation ambient 0 to 50°C, less than 80%rh without condensing
    Power requirement 9-volt battery x1
    Power consumption 37 mA max. if both laser and backlight are turned on
    Battery life at room temperature Approx. 5 hours if both laser and backlight are on;
    12 hours if only the laser is on;
    8 hours if only the backlight is on; or
    55 hours if both laser and backlight are off
    Materials ABS resin
    Dimensions 45 (W) x 157 (H) x 80 (D) mm
    Weight 151 g including a battery
    Standard accessories Hand strap x1, vinyl pouch x1, 9-volt zinc-carbon battery x1, instruction manual